Description
The SFP Services 1250V is a high performance sodium flame photometer designed for filter testing and many other specialized applications. The unit features extraordinary sensitivity and down to 10ng/m3 and when combined with certain accessories can allow the end user to test to the latest ISO 16900 international standards. Sodium flame photometers are in fact specified in certain global standards as the primary measurement method.
Purpose
- To measure the concentration of a standard Na.Cl. aerosol both before and after it has passed through an air filtration device.
Sensitivity
- Better than 10ng/m³ giving a minimum detectable penetration of less than 0.0005%. With a challenge aerosol of 13mg/m³.
Maximum Concentration
- Continuous indication of mass concentrations up to 13mg/m³, but for short periods up to 20mg/m³ if calibrated to this level.
Sample Flow Rate
- 2 l / min +/- 0.1 l / min
Gas Supply
- Hydrogen 0.6 l / m @ 2 bar. [when compressed gas cylinders are used a suitable 2-stage regulator is required with the maximum out put pressure of 5 bar.]
Air Supply
- Minimum flow 50 l/min @ 5 bar.
Electrical Power
- 110 to 264 Vac. 40 to 400 Hz. Single Phase.
Dimensions
- 0,58m long x 0,5m high x 0,29m deep.
Weight
- 24 Kgs
Finish
- Case : – GRP color Mid Grey.
- Front Panel :- Semi-Gloss White.